Coliy Group
- Range: ±100μT ~ ±2000μT
- DC Basic Accuracy: 0.5%
- AC Basic Accuracy: 1%
- Finest resolution: 0.1nT
- DC/AC measurement mode switch
- Frequency response: DC- 2kHz
- Transverse and axial probe optional
GF601 Single-AxisFluxgate Magnetometer
Description:
The 2025-Version Three-axis Fluxgate Magnetometer GF601, madeby COLIY, has the excellent ergonomic design with the latest electronic technology, and it’s one of best choices to measure low-intensity magnetic field. This handheld fluxgate magnetometer can be used in geomagnetism detection, wrapped detection, traffic monitoring, measuring residual magnetism, low-intensity magnetic field measurement, etc.
The key feature of the 2025-Version Magnetometer GF601 is that the size of the probe is reduced to only 24x24x50mm, which is only 1/4 ofthe size of the old version, and it is one of the smallest high-performance fluxgate probes at present, which can be measured in a small space. Magnetometer GF601 adopts a high-precision 24-bit ADC chip, which greatly improves the magnetic field resolution. High performance circuit supports the measurement resolution of 0.1nT within the full measurement range. The power consumption of the whole machine is reduced by 1/4. Magnetometer GF601 boasts with an industrial class 3.2 inches touch panel display that enables customers’ instantaneous and simultaneous measurement results [Max/ Min/ Peak/ Hold/ Alarm/ Polarity, Magnetic Flux Density,and Trend Graph etc.].
Magnetometer GF601 can be used to detect DC magnetic field in DC measurement mode and AC magnetic field in AC measurement mode respectively: GF601 allows DC measurements with a basic accuracy of 0.5%, resolution of 0.1nT and AC measurement with a basic accuracy of 1% and frequency response range of DC~ 2kHz.
In addition, Magnetometer GF601 has passed the CE certification and EMC(Electromagnetic Compatibility) test.
Parameter Comparison | ||
Product Version | 2025-Version | 2020-Version |
Probe Size | 24x24x50mm | 32x32x107mm |
Sampling ADC bits | 24 bits | 18 bits |
Measurement Resolution | 0.1nT | 1nT |
Frequency Response | DC - 2kHz | DC - 1kHz |
Power Consumption | Reduced by 1/4 | High |
Max Range | 2mT | 1mT |
Probe Cable’s Length | Up to 100m | Up to 10m |
Features | |
l GUI Operation System l 3.2 inches color touch LCD l Full 5 display digits l Max/Min/Hold Function l Display trend graph & Alarm l S or N Polar indication l Transverse and axial probe optional | lRange: ±100μT ~ ±2000μT l DC Basic Accuracy: 0.5% l AC Basic Accuracy: 1% l Finest resolution: 0.1nT l DC/AC measurement mode switch l Frequency response: DC- 2kHz lSmall-size Probe (24x24x50mm) |
GF601 Magnetometer Specification:
Model | GF601 / GF601S | ||
Measurement Specification | |||
Accuracy | DC | ±0.5% Reading± 0.1% FS | |
AC | ±1% Reading ± 0.1% FS [fT ≤ 200Hz] ±2% Reading ± 0.1% FS [200Hz < fT ≤ 500Hz] ±10% Reading ± 0.1% FS [500Hz < fT ≤ 2000Hz] | ||
Range (FS) | ±100μT | ±1000μT | |
Finest Display Resolution | 0.1nT | 0.1nT | |
Measurement Resolution | 0.1nT | 0.5nT | |
Typical Temperature Coefficient ppm/℃ | ±100 | ±200 | |
Frequency Response [fT] | DC – 2kHz | ||
Display Digits | 5 digits (More display digits will be shown in supporting PC software) | ||
Hysteresis | <2nT for exposure to up to 2 x full scale <20nT for exposure to 10mT field within 10s | ||
Front Panel | |||
Screen | 3.2 inches color resistive touch LCD,320x240 Pixel | ||
Units | Gauss(G), Tesla(T), Amperes per meter (A/m) | ||
Display Update Rate | 3 readings/second | ||
Display Mode | DC, AC, MAX, MIN, Trend Graph, Alarm, Polarity Indication, etc. | ||
Probe | |||
Start-up Time | 150 ms | ||
Warm-up Time | 15 min | ||
Probe Range | Could be customized within ±2000μT | ||
Probe Cable Length | Standard 1.5m; Customizable longest length of 100m. | ||
USB Interface | |||
Function | 1, Communication Interface: to connect PC with Magnetometer host for monitoring the measurement; 2, Charging interface: to connect PC or mobile power with Magnetometer host for charging | ||
Software/ Driver | SMART PC Software without any driver | ||
Host Specifications | |||
Operating Temperature | +15℃ to +35℃(Rated Accuracy) -10℃ to +60℃(Reduced accuracy) | ||
Storage Temperature | - 20°C to +75°C | ||
Ambient Magnetic Field | <100G (10mT) | ||
Battery | Rechargeable 4500mAH Li-ion | ||
Operating Battery Life | 6 hours; Can be charged by AC power or portable battery | ||
Dimension | 238 mm W × 95 mm H × 42 mm D | ||
Weight | 350g | ||
Certification | CE Certification, EMC Certification |
Note:
1, AC Range mentioned above refers to AC magnetic field measurement range after DC magnetic field is shielded, so if DC magnetic field is not shielded, and users want to measure magnetic field in AC measurement mode, AC Range is equivalent to the value that 0.7 times of DC Range minus DC magnetic field intensity. [SQRT(2)/2 ≈ 0.7]
Description of Magnetometer Selection
GF601 | -A | 10 |
Magnetometer Model
GF601 | Probe Type
-A – Axial Probe -T – Transverse Probe | Measurement Range 01 – ±100μT 02 – ±200μT 05 – ±500μT 10 – ±1000μT 15 – ±1500μT 20 – ±2000μT |
For example:
Model | Descriptions |
GF601-A01 | 3-Axis Fluxgate Magnetometer l With an Axial Fluxgate Probe (±100μT range) |
GF601-A10 | 3-Axis Fluxgate Magnetometer l With an Axial Fluxgate Probe (±1000μT range) |
GF601-T01 | 3-Axis Fluxgate Magnetometer l With a Transverse Fluxgate Probe (±100μT range) |
GF601-T10 | 3-Axis Fluxgate Magnetometer l With a Transverse Fluxgate Probe (±1000μT range) |
Note:
1,Probes can be customized within ±2000μT; Unit Conversion: 1G=100μT=100,000nT;
2,The probe cable length is 1.5m default, and could be customized within 100m.
3,Fluxgate probe housing is made of POM and IP68 protection.
Optional Accessories
Model | Descriptions |
GHOLD100 | 3-Direction Precision Movement Platform, is made of non-magnetic material. Users fixed the probe on the bracket front-end, and then manually rotate the knob so that the probe moves stably along the X, Y, Z-axis to a certain position and lock fixed. Maximum strokes of XYZ axis are 180mm,180mm and 280mm, and positioning accuracy is 0.1mm; center load: 20kg; weight: 6.2kg |
SAMRT PC Software | PC SOFTWARE for Magnetometer |
Probe Extension Cable | MAX length is 100m |